Title: Laboratory study on intracranial hypotension created by pumping the chamber of a hydrocephalus shunt
Authors: Bromby, Adam
Czosnyka, Zofia
Allin, David M
Richards, Hugh K
Pickard, John D
Czosnyka, Marek
Issue Date: 26-Mar-2007
Citation: Fluids and Barriers of the CNS 2007, 4:2
Abstract: Abstract Background It has been reported that pumping a shunt in situ may precipitate a proximal occlusion, and/or lead to ventricular over-drainage, particularly in the context of small ventricles. In the laboratory we measured the effect of pumping the pre-chamber of hydrocephalus shunts on intracranial hypotension. Materials and methods A simple physical model of the CSF space in a hydrocephalic patient was constructed with appropriate compliance, CSF production and circulation. This was used to test eleven different hydrocephalus shunts. The lowest pressure obtained, the number of pumps needed to reach this pressure, and the maximum pressure change with a single pump, were recorded. Results All models were able to produce negative pressures ranging from -11.5 mmHg (Orbis-Sigma valve) to -233.1 mmHg (Sinu-Shunt). The number of pumps required reaching these levels ranged from 21 (PS Medical LP Reservoir) to 315 (Codman Hakim-Programmable). The maximum pressure change per pump ranged from 0.39 mmHg (Orbis-Sigma valve) to 23.1 (PS Medical LP Reservoir). Conclusion Patients, carers and professionals should be warned that 'pumping' a shunt's pre-chamber may cause a large change in intracranial pressure and predispose the patient to ventricular catheter obstruction or other complications.
Description: RIGHTS : This article is licensed under the BioMed Central licence at http://www.biomedcentral.com/about/license which is similar to the 'Creative Commons Attribution Licence'. In brief you may : copy, distribute, and display the work; make derivative works; or make commercial use of the work - under the following conditions: the original author must be given credit; for any reuse or distribution, it must be made clear to others what the license terms of this work are.
URI: http://www.dspace.cam.ac.uk/handle/1810/238039
http://dx.doi.org/10.1186/1743-8454-4-2
Appears in Collections:Scholarly works - Clinical Neurosciences

Files in This Item:

File Description SizeFormat
1743-8454-4-2.xml49.92 kBXMLView/Open
1743-8454-4-2.pdf288.87 kBAdobe PDFThumbnail
View/Open
Additional resources for this item
search for alternative versions in eresources@cambridge
retrieve citation metadata in EndNote format

This item has been accessed 263 times.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.